Specifications

-------------------
Length x width x height - × - × -
Weight -
Worked hours
Hours under power
State excellent
At local norms ---------
Status visible

Description

UPLC

Manufacturer: WATERS
Model: Acquity H-Class
Year: 2015
Dwell volume: <400 μL
Integrated leak management system: leak sensors and leakage safety
Adjustable flow rate: 0.010 to 2.2 mL / min
Cycle time: <30s inject-to-inject
UPLC equipped with:
- QSM module
- SM-FTN Module
- CMP module
- Column heating module
- PDA detector
- Solvent tank
 

QSM module: Quaternary solvent manager
- Model: UPQSM
- Serial number: E14QSM530A
- Number of solvents: mixture combinations up to 4 solvents
- Solvent degassing: integrated vacuum degassing, 4 chambers
- Solvent mixture: automated, pH on-line, ionic strength
- Gradient formation: low-pressure mixture, quaternary gradient
- Main valve: intelligent supply valve
- Flow accuracy ± 1.0% from 0.5 to 2.0 mL / min, 600 to 1000 psi pressure with degassed H2O
- Background noise: <1.0 mAu
- Precision composition accuracy: ± 0.5% absolute from 5% to 90% from 0.5 to 2.0 mL / min
- Compressibility compensation: automatic and continuous
- Priming: Moist, can work up to 4 mL / min
- Pump seals cleaning: equipped with an automated washing system to clean the back of the high-pressure seal and the plunger
- Rate of change of flow:
o Range: 0.01 to 30.00 min to reach 2.0 mL / min
o Default: 0.45 min to reach 2.0 mL / min
- Power: 100 / 240VAC, 50 / 60Hz, 360VA
 
FTN Model: FTN Sample Manager
- Model: UPSMFTN
- Serial number: D14SDI337G
- Injection volume range: 0.1 to 10.0 μL with 0.1 μL increments
- Suction Accuracy: ± 0.2 μL
- Linearity:> 0.999 for standard syringe
- Accuracy:
o <1% RSD 0.2 to 1.9 μL
o <0.5% RSD 2.0 to 10.0 μL
- Number of samples per plate:
o Microtiter trays 96 and 384
o Bottle trays 2.00 mL 48 positions
o Microcentric trays tubes 0.65 mL 48 positions
o Microcentric trays tubes 1.50mL 24 positions
- Sample compartment: 4.0 to 40.0 ° C, adjustable in increments of 0.1 ° C
- Temperature accuracy: ± 0.5 ° C
- Temperature stability: ± 1.0 ° C
- Injection syringe wash: integral, active, programmable
- Minimum sample required: 3 μL residual
- Advanced sample management: auto-dilution, auto-addition and pre-loading
- Power: 100 / 240VAC, 50 / 60Hz, 400VA
 
Column compartment module: CMP
- Model: UPCMA
- Serial number: D14CMP317G
- Reversing valves: double-injector type, seven ports, six positions, reversing valves, left and right, waste and bypass positions
- Column compartment: lodge up to 4 columns in parallel
- Column size: 20 to 150mm long x 2.1 to 4.6mm internal diameter
- Column temperature control: 10.0 to 90.0 ° C, adjustable in increments of 0.1 ° C
- Tracking column
- Service temperature range compartment: from 15 ° C below ambient temperature to 90 ° C
- Precision temperature measurement compartment: ± 1.0 ° C
- Power: 100 / 240VAC, 50 / 60Hz, 400VA
 
 

Column heater:
- Model: CH-30A
- Column size: from 20 to 300mm long x 2.1 to 7.8 mm inside diameter, plus either a guard column or an integrated filter (max 20 mm)
- Column temperature range: 4.0 to 65.0 ° C, in increments of 0.1 ° C
- Measurement accuracy temperature compartment: ± 0.8 ° C
- Dimensions (W x D x H): 203 x 356 x 552 mm
 

PDA detector:
- Model: UPPDATC
- Serial number: E14UPD205A
- Wavelength range: 190 to 500 nm
- Wavelength accuracy: ± 1 nm
- Linearity range: deviation to 2.0 AU 5%
- Optical resolution: 1.2 nm
- Digital resolution: 0.6nm / pixel at 500nm
- Drift: 1.0 x 10-3 / AU / hour / ° C
- Light source: preallanged, Deuterium intelligent technology lamp
- Length of passage: 10mm
- Flow cell volume: 500 nL
- Pressure limit: 1000 psi
- Dimensions (W x D x H): 343 x 607 x 194 mm
- Power: 100 / 240VAC, 50 / 60Hz, 185VA
 